Description
Product Overview
PF0511-M03 is bare gold on a Zerodur disc: no dielectric layer, nothing deposited on top of the metal, 97% average reflectance from 800 nm to 20 µm. It is one percentage point brighter than the protected gold version on the identical substrate, and it is the only coating in the family that combines the highest reflectance with an undisturbed polarization state. The body is 12.7 mm across, 6.0 mm thick, figured to lambda/10 peak to valley at 632.8 nm, polished to 40-20 scratch-dig, parallel to 3 arcmin, 90% clear aperture, part number engraved on the back.
The question worth answering is not what the coating does but what it costs. Thorlabs states that the overcoat on its protected mirrors produces polarization changes that can be difficult to quantify, and that the overcoat sometimes induces absorption in the mid- to far-infrared. Both problems disappear when the overcoat is removed. So does the ability to clean the surface. The published instruction is unusually blunt for a catalogue: bare gold does not oxidise in air, but it can be easily damaged by fingerprints, aerosols, or the slightest contact with any abrasive material; the optic should be handled only when necessary and always held by the sides; latex gloves or a similar covering should be worn; and no attempt should be made to clean the surface other than blowing off dust with clean, dry air or nitrogen.
Thorlabs goes further and warns that mounts using retaining rings are likely to damage these mirrors. That single sentence reclassifies the part: it is not a general-purpose fold mirror that happens to be delicate, it is an optic for a fixed, sealed, pre-aligned assembly where nothing will ever touch the coated face again. Buyers who want a mirror they can wipe should stop reading and take the protected gold -M01 instead.
Where the trade does pay is narrow and specific. In a strictly polarization-maintaining interferometer, an ellipsometer, or a femtosecond stretcher, an unquantified polarization rotation is not a small error, it is an uncalibratable one, and 1% of extra reflectance across every bounce is real throughput. In the far infrared beyond roughly 7 µm the overcoat absorption of the protected version is also measurable. For those three cases the handling discipline is cheap. For everything else it is not.
Key Features
- 97% average from 800 nm to 20 µm: the highest reflectance of any gold coating Thorlabs puts on this substrate
- No overcoat at all: no dielectric absorption in the mid and far infrared, and no thin-film dispersion to chirp a short pulse
- Polarization left alone: the reason Thorlabs names polarization-sensitive and ultrafast work as the target applications for this coating
- Zerodur body: 0 plus or minus 0.100 x 10-6/degC, so the low-drift argument and the polarization argument can be had in the same part
- Blow-clean only: clean dry air or nitrogen, nothing else, ever
- Laser rated at 10.6 µm: 4 J/cm2 for 100 ns pulses and 1000 W/cm CW, the highest CW figure in the metallic Zerodur range
Technical Specifications
| Parameter | Value |
|---|---|
| Item Number | PF0511-M03 |
| Coating | Unprotected Gold (bare gold, no overcoat) |
| Wavelength Range | 800 nm to 20 µm |
| Average Reflectance | 97% from 800 nm to 20 µm |
| Angle of Incidence | 0 to 45 degrees |
| Substrate | Zerodur (Schott registered trademark, substrate grade) |
| Diameter | 12.7 mm (1/2 inch), +0.0 / -0.1 mm |
| Thickness | 6.0 mm (0.24 inch), plus or minus 0.2 mm |
| Clear Aperture | 90% of the diameter |
| Front Surface Flatness | lambda/10 peak to valley at 632.8 nm |
| Surface Quality | 40-20 scratch-dig |
| Parallelism | 3 arcmin maximum |
| Coefficient of Thermal Expansion | 0 plus or minus 0.100 x 10-6/degC |
| Back Surface | Fine ground, item number engraved |
| Permitted Cleaning | Blowing off dust with clean, dry air or nitrogen only |
| Pulsed Damage Threshold | 4 J/cm2 at 10.6 µm, 100 ns, 1 Hz, over 0.435 mm |
| CW Damage Threshold | 1000 W/cm at 10.6 µm over 1.18 mm |
| Source | Thorlabs official series specification table |
Note the shape of those two laser numbers. Both are quoted at 10.6 µm, whereas the protected gold ratings are quoted at 1064 nm for the pulsed case. They are not a like-for-like ranking of the two coatings; they are the data Thorlabs has. What they do show is that at CO2 wavelengths this film carries the highest CW rating in the family.
Applications
- Polarization-maintaining interferometers: where a coating-induced rotation that nobody can quote a number for is worse than a known 1% loss
- Ellipsometry and polarimetry: instruments whose entire output is a polarization state, and which therefore cannot tolerate a film that perturbs it at oblique incidence
- Femtosecond and ultrafast systems: no overcoat means no group-delay dispersion added by a coating, and no chirp to pre-compensate
- Far-infrared benches: beyond about 7 µm the overcoat absorption of the protected version becomes visible in the throughput budget
- Sealed CO2 heads: fixed optics inside an enclosure, never touched, rated 1000 W/cm CW at 10.6 µm
Selection Guide
This part is chosen by elimination, not by preference. Answer three questions; if any answer is no, buy the protected version.
- Is the polarization state a specified output of your system? If polarization is merely incidental, the overcoat of -M01 is not costing you anything and the cleanability is free.
- Will anything ever touch the coated face after installation? Retaining rings, dust caps that press on the surface, and routine cleaning all disqualify this part. Thorlabs explicitly warns that retaining-ring mounts are likely to damage it.
- Is the assembly sealed and pre-aligned? If the optic goes in once and the enclosure closes, the handling risk is a one-time event that trained hands can manage.
- Then, and only then, is the 1% worth it. If you have answered yes three times, 97% with clean polarization is the best combination available on this substrate.
Series Comparison
The comparison that matters for this part is not reflectance, it is what the overcoat does to the light. The table below separates the two effects.
| Attribute | Protected Gold -M01 | Unprotected Gold -M03 |
|---|---|---|
| Average reflectance, 800 nm to 20 µm | 96% | 97% |
| Overcoat | Dielectric protective layer | None |
| Polarization at oblique incidence | Changed, and hard to quantify | Essentially undisturbed |
| Dispersion added by the film | Yes | No |
| Absorption in the mid to far infrared | Present, induced by the overcoat | Absent |
| Cleaning | Solvent drag wipe permitted | Blowing off dust only |
| Retaining-ring mounts | Acceptable with a soft ring | Likely to damage the surface |
| Intended environment | Open bench, industrial machine | Sealed, fixed, pre-aligned assembly |
FAQ
Can I clean it if it gets dusty? Only by blowing. Thorlabs permits clean, dry air or nitrogen and nothing else, and adds that any other cleaning method may damage the surface. There is no solvent wipe, no lens tissue, no first-contact polymer peel and no ultrasonic bath that is sanctioned for bare gold. If the environment cannot be kept clean enough for blow-only maintenance, this is the wrong coating for the job regardless of how good the numbers look.
What actually kills it, and how fast? Contact, not chemistry. Bare gold does not oxidise in air, so shelf life in a clean container is long. The failure modes Thorlabs lists are fingerprints, aerosols and the slightest contact with any abrasive material, and the practical consequence is that the damage happens the first time somebody reaches for it with a tissue. A part that is installed once and left alone can last for years; a part that is handled weekly will not.
Why do retaining-ring mounts damage it? Because a metal ring bears on the coated face at the rim and any grit trapped between ring and coating is dragged across gold that is softer than the grit. Thorlabs states the risk directly. Use a mount that holds the optic by its edge or its barrel, or fit a soft interface, and accept that the clamping force has to be low enough not to distort a 6 mm disc anyway.
Is 1% really worth losing the ability to clean it? Only where the second property is the real purchase. If you need polarization fidelity, dispersion-free reflection or far-infrared throughput, the 1% is a bonus and the handling rule is the price of the thing you actually wanted. If you need none of those, the protected coating is better value and far more forgiving.
Usage Precautions
- Handle only when necessary, always by the sides, with latex gloves or an equivalent barrier. Finger oil reaching the surface is permanent damage, not a smudge.
- Never wipe. Blow off dust with clean, dry air or nitrogen, inspect, and stop there.
- Do not use a retaining-ring mount, and do not let a dust cap press on the coated face.
- Fit and align the optic before the rest of the assembly is closed up; every subsequent touch is a risk with no upside.
- Check the beam stays inside the 90% clear aperture, because a hot spot at the edge of a bare gold film has no overcoat to protect it.
Disclaimer
Specifications on this page are transcribed from Thorlabs official public documentation and are provided for selection reference only. The current official specification sheet governs. Thorlabs may change specifications, coatings and appearance without notice. Prices, availability and delivery times are confirmed by our sales team at the time of order; nothing on this page is a binding offer.
Contact Us
Gaofei Laser supplies genuine Thorlabs stock and can quote this item with current availability, lead time and volume pricing. Send the item number, the quantity you need and the wavelength you are working at, and we will confirm the coating, the substrate and the mounting interface before you order.
- Sales hotline: +86 159 2722 0467
- Email: 859818709@qq.com
- Website: www.soar-laser.cn









Reviews