Thorlabs PFA52505 Active Isolation Breadboard Support Frame, 800 mm Height, 750 x 900 mm

Thorlabs PFA52505 Active Isolation Breadboard Support Frame, 800 mm Height, 750 x 900 mm

Price on request
Category:

Description

Product Overview

PFA52505 is an active isolation breadboard support frame for a 750 x 900 mm (30 in x 36 in) breadboard, built on an 800 mm (31.5 in) high frame. It carries 700 kg (1,545 lbs) on four independent, air mounted, active vibration isolators, levels itself pneumatically to ±0.25 mm (0.010 in) repeatability, and holds a vertical resonant frequency below 1.6 Hz and a horizontal resonant frequency below 1.2 Hz. At resonance the vertical transmissibility is below 13 dB and the horizontal transmissibility is below 8 dB. At 5 Hz both axes are at -18 dB, which is 87.5 percent of the ground motion removed, and at 10 Hz both axes are at -34 dB, which is 98 percent removed. The construction is all steel with a Black Paint finish, the air supply requirement is 690 kPa (100 psi) maximum, and the PWA061 caster set is available as an option.

The reason to read this page rather than the page for the 700 mm version of the same outline is height, and the first thing to get straight about height is that 800 mm is not the height of the working surface. The frame is the thing that stands on the floor. The breadboard is a separate item that bolts on top of it, and the breadboard has its own thickness, which varies by a factor of more than four across the boards normally used on a bench this size. A 60 mm Nexus breadboard puts the working surface at 860 mm. A 110 mm Nexus breadboard puts it at 910 mm. An internally damped 58 mm board puts it at 858 mm, a 25 mm aluminium board at 825 mm and a 55 mm aluminium board at 855 mm. The 700 mm frame produces exactly the same five surfaces, each 100 mm lower.

The second thing to get straight is that the surface height is not the working height either. Nobody adjusts an optic by reaching down to the table surface. The work happens at the optical axis, which is the surface plus whatever the mount contributes. A 12.7 mm (0.5 in) post carrying a kinematic mount puts the axis about 100 mm above the surface. A 25 mm (1 in) post carrying a larger mount puts it about 150 mm above. On this frame the axis therefore lands between 925 mm and 1060 mm above the floor, depending on which breadboard and which mount are fitted. On the 700 mm version the same combinations land between 825 mm and 960 mm. That 100 mm shift is the entire subject of this page, and it is worth more than it sounds.

For a standing operator the two reference points are the elbow and the eye. In the Chinese adult population described by GB/T 10000-1988 the standing elbow height is about 1024 mm for men and about 960 mm for women, and the standing eye height is about 1568 mm and 1454 mm respectively. Fine optical work – turning a differential adjuster, checking a focus, seating a fiber in a connector – is normally performed with the hands slightly above elbow height, in a band running roughly 50 to 100 mm above it. That gives a fine-work band of about 1074 to 1124 mm for a male operator and about 1010 to 1060 mm for a female operator, so the band that serves both is roughly 1010 to 1124 mm.

Put the two sets of numbers together and the result is not what the marketing language would suggest. A surface height of 825 to 910 mm is below the fine-work band in every case, on the 800 mm frame as much as on the 700 mm one. What the 800 mm frame actually does is put the optical axis – not the surface – inside that band. On the 700 mm frame the axis sits at 825 to 960 mm, which is at or below the bottom of the band for a female operator and well below it for a male operator; the hands end up working below the elbow, the shoulders drop and the upper back rounds. On this frame the axis sits at 925 to 1060 mm, which straddles the band. This is why the recommendation for a standing alignment station is not to raise the table but to choose the frame that puts the beam where the hands want to be.

The 750 x 900 mm outline is the smallest rectangle in the family that a single operator can use as a full station rather than as a shelf. Its area is 0.675 m2, its perimeter 3300 mm, its diagonal 1171.5 mm, and its aspect ratio 1 to 1.2, which leaves a clear span of about 700 mm along the 900 mm axis once 100 mm at each end is reserved for mounts and beam clearance. Facing the 900 mm side, an operator has a 900 mm wide by 750 mm deep surface. The maximum forward reach of a 50th percentile male is about 834 mm, but the comfortable working reach is about 65 percent of that, around 542 mm, and standing 100 mm off the front edge costs another 100 mm. That leaves a comfortable band of about 442 mm at the front of the 750 mm depth, with the rear 308 mm available only by leaning over the bench. Along the 900 mm width, comfortable coverage at a working depth of 300 mm is about 650 mm, so the station needs two stances rather than one.

None of the isolation performance changes with the outline or the height. The isolators are the same components on every frame in the PFA range, so the resonant frequencies, the transmissibility figures, the 700 kg rating, the ±0.25 mm leveling repeatability and the 690 kPa air requirement are identical from the 750 x 750 mm frame to the 1200 x 1500 mm frame. What this size adds is a usable single-operator surface at a standing working height, in a footprint that will go through a standard door and does not need a dedicated room.

The sections below deal with the three separate vertical numbers that this frame is often confused over, with the geometry of the 750 x 900 mm surface, with the vibration environment the frame has to work in, and with what changes and what does not when the 800 mm frame is compared with its 700 mm twin. Anyone who already knows they want a 900 mm high working surface and only needs the isolation data can read the specification table and stop there.

Key Features

  • 800 mm (31.5 in) frame height, which produces a working surface of 825 to 910 mm once a 25 to 110 mm breadboard is fitted, and an optical axis of roughly 925 to 1060 mm above the floor.
  • 750 x 900 mm (30 in x 36 in) outline: 0.675 m2 of surface, 3300 mm of perimeter, a 1171.5 mm diagonal and a 1 to 1.2 aspect ratio that leaves about 700 mm of clear span along the 900 mm axis.
  • Vertical resonant frequency below 1.6 Hz and horizontal resonant frequency below 1.2 Hz, which is roughly half the 3.5 Hz corner of a passive isolation frame.
  • Vertical transmissibility below 13 dB and horizontal transmissibility below 8 dB at resonance, against 24 dB for a standard passive frame, which amplifies rather than isolates at its corner.
  • Vertical and horizontal transmissibility of -18 dB at 5 Hz, equivalent to 87.5 percent of the ground motion removed on both axes.
  • Vertical and horizontal transmissibility of -34 dB at 10 Hz, equivalent to 98 percent removed, which is 18 dB better than the passive figure at the same frequency.
  • 700 kg (1,545 lbs) load capacity, which is a total that includes the breadboard and not a payload figure.
  • Dual chamber, damped pneumatic spring in the vertical direction: the breadboard is supported on the pressure in the air chamber, sits on a piston sealed to the upper chamber by a rolling rubber diaphragm, and is damped by air forced through a high damping orifice between the two chambers.
  • Trifilar suspension system in the horizontal direction: the pneumatic isolators stand on a three wire pendulum whose restoring force comes from gravity, with oil free damping to suppress the horizontal resonance.
  • Pneumatic self-leveling through precision three-way valves driven by the motion of the working surface, returning to the original level within ±0.25 mm (0.010 in) and compensating automatically for changes in load distribution.
  • Isolator height adjustment of ±6 mm (0.25 in) and leg height adjustment of ±10 mm (0.4 in), giving 26 mm (1.02 in) of total system travel for setting the working height.
  • 690 kPa (100 psi) maximum air pressure, supplied continuously; when the air supply is removed the breadboard rests securely on top of the frame with the isolation system disabled.
  • All steel construction with a Black Paint finish, which gives a stiff, thermally stable frame that does not creep under a static load.
  • Four independent isolators rather than three, so the frame remains stable when a load is placed off centre and so that a single isolator can be serviced without emptying the bench.
  • PWA061 caster set available as an option: 360 degree swivel, high durability non marking polyurethane wheels, 200 kg (440 lbs) per caster and 800 kg (1,760 lbs) for the set of four, which lets the frame be moved with the equipment still mounted.

Technical Specifications

The table below reproduces the manufacturer data for this model. The isolation figures are properties of the isolator design and are therefore common to every outline and both frame heights in the PFA range; the rows that vary between models are the breadboard size, the frame height and the geometry derived from the outline.

Specification Value
Model PFA52505
Product class Active isolation breadboard support frame
Breadboard size accepted 750 x 900 mm (30 in x 36 in)
Frame height 800 mm (31.5 in)
Vertical resonant frequency below 1.6 Hz
Horizontal resonant frequency below 1.2 Hz
Vertical transmissibility at resonance below 13 dB
Horizontal transmissibility at resonance below 8 dB
Vertical transmissibility at 5 Hz -18 dB (87.5% isolation)
Horizontal transmissibility at 5 Hz -18 dB (87.5% isolation)
Vertical transmissibility at 10 Hz -34 dB (98% isolation)
Horizontal transmissibility at 10 Hz -34 dB (98% isolation)
Load capacity 700 kg (1,545 lbs)
Isolators height adjustment range ±6 mm (0.25 in)
Self leveling repeatability ±0.25 mm (0.010 in)
Legs height adjustment range ±10 mm (0.4 in)
Air pressure required 690 kPa (100 psi) maximum
Finish Black Paint
Construction All steel
Isolator count and type Four independent, air mounted, active vibration isolators
Leveling method Pneumatic self-leveling, precision three-way valves
Optional casters PWA061 (set of 4)
Largest breadboard supported by family 1200 x 1800 mm (48 in x 72 in)
Outline area 0.6750 m2
Outline perimeter 3300 mm
Outline diagonal 1171.5 mm
Aspect ratio 1 to 1.2, long axis 900 mm
Clear span along the 900 mm axis about 700 mm after 100 mm at each end for mounts

Table 1. Manufacturer specification for PFA52505, with outline geometry derived from the stated breadboard size. Values are reproduced as published and are not rounded or converted.

Two notes on the height adjustment rows. The ±6 mm and ±10 mm figures above are the ones given in the manufacturer specification table and they refer to two different components: the isolators themselves have 6 mm of travel each way, and the leveling feet at the bottom of the legs have 10 mm each way. The manufacturer product description separately quotes leveling pads at ±9.5 mm (±0.375 in) and leveling feet at ±5 mm (±0.2 in). Those are different parts described in a different place, not a contradiction in the specification, and for selection purposes the specification table above governs. The separate figure of 26 mm (1.02 in) that appears in the manufacturer text is the total height range of the assembled system, not a third adjustment.

The second note concerns the transmissibility figures and the percentages printed beside them. A transmissibility of -34 dB means the motion transmitted to the breadboard is 10 raised to the power of minus 34 divided by 20, which is a factor of 0.0200. In other words 2.0 percent of the ground motion gets through and 98 percent is removed: the 98 percent is the isolation, not the transmission. Likewise -18 dB is a factor of 0.1259, so 12.6 percent gets through and 87.4 percent is removed, which the manufacturer rounds to 87.5 percent. Reading the brackets the other way round is the single most common mistake made with these numbers. The dB value is negative because it describes attenuation; the percentage in brackets is the fraction attenuated, so a larger percentage is better, and 98 percent is better than 87.5 percent.

The vertical isolation principle is a dual chamber, damped pneumatic spring. The breadboard is supported by the pressure in an air chamber and sits on a piston. Between the piston and the upper chamber the seal is a rolling rubber diaphragm, which allows almost frictionless vertical motion while preventing horizontal displacement. When the floor or the bench top moves, the air is forced through a high damping orifice from the lower chamber into the upper chamber, which damps the oscillation and shortens the settling time substantially. The volumes of the two chambers are chosen so that the damping is maximised while the resonant frequency stays low, which is why this design reaches 1.6 Hz where a simple air bag would either be too stiff or too underdamped.

The horizontal isolation principle is a trifilar suspension system. The pneumatic vertical isolators sit on a three wire pendulum, and the restoring force after a horizontal displacement comes from gravity rather than from a spring. Oscillation at the horizontal resonant frequency is suppressed by oil free vibration absorbing dampers connecting the bottom of each vertical isolator to the outer cylinder. Because the restoring force is gravitational, the horizontal corner frequency is set by the effective pendulum length rather than by a spring rate, and the 1.2 Hz figure implies an effective length of about 173 mm, which is short enough to keep the assembly compact and long enough to put the corner well below the 3.5 Hz of a passive frame.

Leveling is handled by precision three-way valves that form a self-leveling system. The valves are actuated by movement of the working surface itself, so a disturbance that tilts the breadboard opens the appropriate valve, admits or exhausts air, and returns the surface to within ±0.25 mm of its original level. The same mechanism compensates automatically when equipment is added to or removed from one end of the bench, which is the behaviour that makes a four isolator frame practical for a laboratory where the load changes from week to week. When the air supply is removed, the breadboard rests securely on top of the frame with the isolation system disabled, so the bench becomes a rigid table rather than becoming unstable.

The table below converts the frame height into an actual working surface height for the breadboard thicknesses in normal use, and repeats the calculation for the 700 mm version of the same outline so that the 100 mm difference is visible rather than assumed.

Breadboard fitted Thickness Surface on this 800 mm frame Surface on the 700 mm version
Nexus breadboard 60 mm (2.36 in) 860 mm (33.9 in) 760 mm (29.9 in)
Nexus breadboard, thick 110 mm (4.33 in) 910 mm (35.8 in) 810 mm (31.9 in)
Internally damped breadboard 58 mm (2.28 in) 858 mm (33.8 in) 758 mm (29.8 in)
Aluminium breadboard, thin 25 mm (0.98 in) 825 mm (32.5 in) 725 mm (28.5 in)
Aluminium breadboard, standard 55 mm (2.17 in) 855 mm (33.7 in) 755 mm (29.7 in)

Table 2. Working surface height, calculated as frame height plus breadboard thickness. The 800 mm frame adds 100 mm to every row; it does not change any other number in this table.

The next table adds the mount, because the surface is where things are bolted down and the axis is where the work happens. Three mount heights are shown because the choice between a short post and a tall one moves the axis by as much as the choice between the two frames does, which is a useful reminder that the frame is only one of the two variables.

Breadboard and mount Axis on this 800 mm frame Axis on the 700 mm version Comment
60 mm board, 100 mm mount centreline 960 mm 860 mm Short post, small optic
60 mm board, 125 mm mount centreline 985 mm 885 mm Typical 0.5 in kinematic mount
60 mm board, 150 mm mount centreline 1010 mm 910 mm 1 in post, larger optic
110 mm board, 100 mm mount centreline 1010 mm 910 mm Stiff board, short post
110 mm board, 125 mm mount centreline 1035 mm 935 mm Stiff board, typical mount
110 mm board, 150 mm mount centreline 1060 mm 960 mm Stiff board, tall post

Table 3. Optical axis height above the floor, calculated as frame height plus breadboard thickness plus mount centreline height. Values are arithmetic, not manufacturer data.

Read Table 3 against the fine-work band of roughly 1010 to 1124 mm derived in the overview and the usefulness of the 800 mm frame becomes concrete. On the 700 mm version, only the stiffest combination in the table – a 110 mm board with a 150 mm mount – reaches 960 mm, and even that is below the band. On this frame, four of the six combinations land inside the band and the remaining two are within 50 mm of it. If the station is to be used standing for hours at a time, that is the difference between an operator who can work with relaxed shoulders and one who cannot, and it costs 100 mm of frame height rather than a taller bench, a platform or a change of posture.

Applications

The applications below are the ones where a 750 x 900 mm surface at a standing working height is the right answer rather than a compromise. In each case the deciding factor is either the number of people who need to use the bench, which is one, or the height at which the work has to be done, which is standing.

  • Single-operator alignment and assembly stations. One person, one beam path, one shift. A 750 x 900 mm surface holds a laser head, four to six mounts, a translation stage and a detector with room to spare, and the 800 mm frame puts the beam at a height where the operator does not have to stoop to turn an adjuster. This is the most common use of this outline.
  • Fiber coupling and photonics packaging. Coupling a free-space beam into a fiber requires the hands to be at the axis and the eyes to be at the same height as the coupling optics, often for twenty minutes at a time. The axis at 985 mm on a 60 mm board is close to eye level minus a comfortable downward angle, which is exactly the geometry this frame provides.
  • Confocal and single-molecule imaging benches. These need a small, very quiet surface rather than a large one, because the whole experiment fits within 600 mm. The 0.675 m2 outline is enough, the -34 dB at 10 Hz removes the building services noise that dominates in that band, and the four isolators keep the bench stable when a heavy detector is added at one end.
  • Interferometric metrology at standing height. Where a reference arm and a measurement arm have to be adjusted repeatedly while watching fringes, the operator is standing and looking along the bench. A 700 mm span along the 900 mm axis is enough for a folded Michelson or a Twyman-Green, and the low horizontal corner of below 1.2 Hz matters more than the vertical figure because the fringe pattern is most sensitive to lateral drift.
  • Teaching and demonstration laboratories. A bench that one student can operate from one side, that will go through a 900 mm door on its side, and that survives being pushed against a wall between sessions. The PWA061 casters are rated 200 kg each and 800 kg for the set, so the frame can be moved with the breadboard and the optics still mounted.
  • Probe stations and semiconductor test benches. A probe station is operated standing, with the hands on micromanipulators at chest height and the eyes on a microscope. The 800 mm frame plus a 60 mm board gives 860 mm of surface, and the probe platen then sits at about the right height without an extra riser.
  • Small cold atom and quantum optics setups. A vacuum chamber, four or five frequency paths and a detection arm will fit on 0.675 m2 if the layout is drawn first. The horizontal isolation at -18 dB at 5 Hz is the relevant figure here, because the low frequency building motion that shifts a locked laser is exactly what a passive frame cannot remove.

Selection Guide

Choosing between the 700 mm and 800 mm versions of this outline is a two step decision, and the order matters. The first step is to establish how much the experiment cares about low frequency vibration, because that decides whether an active frame is required at all. The second step is to establish the working height, because that decides which of the two frames to order. Doing it the other way round produces a comfortable bench in a room where the experiment will never settle.

The table below is the manufacturer summary of common vibration sources, and it is the reason the first step comes first. Read the frequency column rather than the amplitude column: amplitude tells you how much a source moves, frequency tells you whether the isolation system can do anything about it.

Source Frequency (Hz) Amplitude (inches)
Air compressors 4 – 20 10^-2
Handling equipment 5 – 40 10^-3
Pumps (vacuum, compressed, or non-compressed fluids) 5 – 25 10^-3
Building services 7 – 40 10^-4
Foot traffic 0.5 – 6 10^-5
Acoustics 100 – 10,000 10^-2 to 10^-4
Punch presses up to 20 10^-2 to 10^-5
Transformers 50 – 400 10^-4 to 10^-5
Elevators up to 40 10^-3 to 10^-5
Building motion 46 / height in metres 10^-1
Building pressure waves 1 – 5 10^-5
Railroad 5 – 20 ±0.15 g
Highway traffic 5 – 100 ±0.001 g

Table 4. Common vibration sources and their frequency content, as published by the manufacturer. Ambient noise in a laboratory is typically concentrated between 10 and 100 Hz.

Three rows in that table decide whether an active frame is necessary. Foot traffic runs from 0.5 to 6 Hz. Building pressure waves run from 1 to 5 Hz. Air compressors run from 4 to 20 Hz. All three extend below 2 Hz, and a passive frame with a 3.5 Hz corner cannot isolate anything below its own resonance – at 3.5 Hz it amplifies by 24 dB, which is a factor of about 16. This is the argument for the below 1.6 Hz vertical corner and the below 1.2 Hz horizontal corner: the corner has to be pushed below the lowest source that matters, not merely made lower than it was.

The second step is the height, and the procedure is arithmetic rather than judgemental:

  • Fix the breadboard first, because its thickness is the largest single term after the frame. A 60 mm board on this frame gives 860 mm of surface; a 110 mm board gives 910 mm.
  • Add the mount centreline height for the tallest optic that will sit on the bench, not the shortest. A 150 mm mount on a 60 mm board puts the highest axis at 1010 mm.
  • Compare that axis height with the fine-work band for the people who will actually use the bench. For a mixed team the band is roughly 1010 to 1124 mm; for a team of shorter operators it is roughly 1010 to 1060 mm and the lower end of the band is the one that matters.
  • If the axis lands below 960 mm, the 800 mm frame is the one to order. If it lands above 1100 mm already, the 700 mm frame will do and the extra 100 mm would push the work above the band.
  • Check the seating question separately. A seated operator needs the surface at about 700 to 760 mm, which is the 700 mm frame with a thin board, and no amount of optical argument makes an 860 mm surface comfortable from a chair.
  • Check the air supply. 690 kPa (100 psi) maximum, dry and oil free, at a flow rate sufficient to recover from a load change. A frame this size will not perform to specification on a workshop compressor that cycles between 400 and 800 kPa.
  • Decide on casters. The PWA061 set is rated 200 kg per caster and 800 kg for the four, against a frame rating of 700 kg, so the set is adequate for the full load with about 14 percent in hand, but the floor has to be flat and free of thresholds for the frame to roll without tilting.

One further check is worth making before the order, and it is specific to a small outline. On a 750 mm deep bench the rear 300 mm is outside comfortable reach for a standing operator, so anything that needs adjusting during a session has to be inside the front 450 mm. That is a layout constraint rather than a frame constraint, but it is easier to satisfy on a 900 mm wide bench than on a 1200 mm wide one, because the operator has less width to cover laterally: at a working depth of 300 mm the comfortable lateral coverage is about 650 mm, so two stances cover the 900 mm width with margin.

If the bench is to be used by more than one person at a time, this outline is the wrong one. Two operators facing each other need at least 900 x 1200 mm, and two operators side by side on the same 900 mm side need about 600 mm of width each, which leaves nothing for the optics. The 900 x 1200 mm frame in the same family is the smallest that supports a genuine two person station.

Series Comparison

The PFA family is one isolator design on six outlines and two frame heights, which makes the comparison unusually clean: nothing about the isolation changes when the outline or the height changes. The table below sets out the six 800 mm outlines with the geometry that decides how much work each will take, followed by the pairing with the 700 mm version of the same outline.

Model Breadboard Size Area (m2) Perimeter (mm) Diagonal (mm) Clear Span Typical Use
PFA52509 750 x 750 mm 0.5625 3000 1060.7 about 550 mm Single instrument, smallest footprint
PFA52505 750 x 900 mm 0.6750 3300 1171.5 about 700 mm One operator, one beam path
PFA52508 750 x 1200 mm 0.9000 3900 1415.1 about 1000 mm One operator, long path
PFA52507 900 x 1200 mm 1.0800 4200 1500.0 about 1000 mm Two operators facing each other
PFA52502 900 x 1500 mm 1.3500 4800 1749.3 about 1300 mm Two stations or one long path
PFA52504 1200 x 1500 mm 1.8000 5400 1920.9 about 1300 mm Multi-station, whole room

Table 5. The six 800 mm outlines in the PFA range. Area, perimeter and diagonal are calculated from the stated breadboard size; clear span assumes 100 mm reserved at each end for mounts and beam clearance.

The pairing table is the one to use when the question is whether 100 mm of extra frame height is worth it. The isolation specification, the load rating, the leveling repeatability, the air requirement and the breadboard compatibility are identical between the two columns; the only difference is the frame height and, as a consequence, the surface and axis heights.

Breadboard Size 700 mm Frame 800 mm Frame Surface (60 mm Board) Relative Cost of 800 mm
750 x 750 mm PFA51509 PFA52509 760 mm against 860 mm about 0.4 percent more
750 x 900 mm PFA51505 PFA52505 760 mm against 860 mm about 0.4 percent more
750 x 1200 mm PFA51508 PFA52508 760 mm against 860 mm about 0.4 percent more
900 x 1200 mm PFA51507 PFA52507 760 mm against 860 mm about 0.4 percent more
900 x 1500 mm PFA51502 PFA52502 760 mm against 860 mm about 4.2 percent more
1200 x 1500 mm PFA51504 PFA52504 760 mm against 860 mm about 4.2 percent more

Table 6. The two frame heights for each outline. The relative cost column compares published list prices for the two heights of the same outline and is expressed as a percentage only; no absolute price is stated on this page.

The cost pattern in the last column is worth a sentence because it is not what most buyers expect. On the four smaller outlines the 800 mm frame costs well under half a percent more than the 700 mm frame, which is small enough that the decision can be made entirely on working height. On the two largest outlines the uplift is around four percent, because the longer legs are a different structure rather than the same structure with more material. This outline is in the first group, so the height decision here costs almost nothing.

The outline that this one should be compared with is PFA52508, the 750 x 1200 mm frame, which has the same 750 mm depth and 300 mm more length for 0.225 m2 more area. If the beam path fits within a 700 mm clear span, the shorter bench is the better choice: it costs less, it occupies less floor, and the operator never has to walk. If the path needs more than 700 mm of run before a fold, the extra length pays for itself immediately, because one avoided fold is worth about 3 percent of the light and, on an ultrafast system, one avoided set of dispersive mirrors.

Compared with PFA52507, the 900 x 1200 mm frame, the difference is not length but depth. The 750 mm depth of this frame keeps the far edge within maximum reach, at about 734 mm from the front edge for a 50th percentile male standing 100 mm back, which means the whole surface is reachable without walking round. On the 900 x 1200 mm frame the 900 mm depth puts the far edge beyond reach entirely, so a two person station becomes the natural arrangement. For one operator, the shallower bench is faster to work at even though it holds less.

FAQ

Is the working surface 800 mm above the floor?

No. 800 mm is the height of the frame, which is the top of the structure that the breadboard bolts onto. The working surface is 800 mm plus the breadboard thickness: 860 mm with a 60 mm Nexus board, 910 mm with a 110 mm board, 858 mm with an internally damped 58 mm board, 825 mm with a 25 mm aluminium board and 855 mm with a 55 mm aluminium board. If a drawing specifies a surface height, subtract the breadboard thickness before choosing a frame.

What height is the optical axis on this frame?

Between about 925 mm and 1060 mm above the floor, depending on the breadboard and the mount. The axis is the surface height plus the height of the mount centreline, which is roughly 100 mm for a 12.7 mm post with a small kinematic mount and roughly 150 mm for a 25 mm post with a larger one. The axis, not the surface, is the height at which an operator’s hands work, so it is the number to compare against elbow height.

Why does the horizontal resonant frequency matter as much as the vertical one?

Because lateral ground motion moves a beam sideways, and a beam that moves sideways by a micrometre at the far end of a two metre path is a fringe shift. The horizontal corner on this frame is below 1.2 Hz, which is low enough to sit under foot traffic and building pressure waves, and the horizontal transmissibility at 10 Hz is -34 dB, the same as the vertical figure. A frame that quotes only a vertical number has not told you how it behaves laterally.

What does -34 dB at 10 Hz actually mean in percent?

It means 2.0 percent of the ground motion is transmitted and 98 percent is removed. The percentage in brackets in the specification is the isolation, not the transmission. The same applies at 5 Hz, where -18 dB is 12.6 percent transmitted and 87.4 percent removed, quoted as 87.5 percent. If you want the transmission figure, divide the dB value by 20, raise 10 to that power and multiply by 100.

Can this frame be used without a compressed air supply?

It can be used as a rigid table but not as an isolation frame. With the air supply removed the breadboard rests securely on top of the frame with the isolation system disabled, so nothing is unstable, but the resonant frequencies and transmissibility figures no longer apply. The air requirement is 690 kPa (100 psi) maximum, and it has to be supplied continuously, dry and oil free, because the leveling valves and the damping orifices are small passages that will not tolerate moisture or oil carry over.

How much load can actually be put on the bench?

The 700 kg (1,545 lbs) rating is a total and it includes the breadboard. A 750 x 900 mm, 60 mm Nexus board weighs on the order of 40 to 50 kg, so the equipment budget is about 650 kg before any margin. Applying a 20 to 30 percent margin brings the working payload to roughly 450 to 520 kg. That is a large amount for a bench this size, and the practical limit is more likely to be the floor loading than the frame rating.

Does the frame isolate better if it is more heavily loaded?

Not in the way a steel spring would. In a pneumatic isolator the air spring stiffness rises with the internal pressure, and the pressure rises with the load, so the ratio of stiffness to mass and therefore the resonant frequency stays close to the quoted value across the working load range. That is one of the advantages of an air spring over a coil spring, where the frequency would climb as the square root of the inverse of the mass as the load came off. The isolator does have a working load window, however, and at a very light load the piston approaches the top of its stroke and leveling authority is lost.

Can I move the frame with the equipment still on it?

The PWA061 caster set is intended for exactly that. Each caster is rated 200 kg (440 lbs) and the set of four is rated 800 kg (1,760 lbs), against a frame rating of 700 kg, so the casters are adequate for the full load with about 14 percent in hand. The constraint is the floor: the frame must not be rolled over thresholds, cable ducts or steps, because a tilt while rolling puts the whole load onto two casters and, more importantly, will spill anything that is not clamped.

What is the difference between the ±6 mm, ±10 mm and 26 mm height figures?

They refer to three different things. The ±6 mm is the travel of the isolators themselves. The ±10 mm is the travel of the leveling feet at the bottom of the legs. The 26 mm (1.02 in) quoted in the manufacturer description is the total height range of the assembled system, which is what you can use to trim the working height on site. None of them is a substitute for the frame height: 100 mm of difference between the two versions of this outline cannot be absorbed by any of these adjustments.

Will a 750 x 900 mm breadboard fit on a 700 mm frame of the same outline?

Yes, and the reverse also holds: the breadboard compatibility is identical between the two frame heights, because the mounting interface is the same. The only thing that changes is where the surface ends up. That is worth knowing if a laboratory already owns the 700 mm frame and is considering a second bench, because breadboards and accessories can be shared between the two.

How much space should be left around the frame?

At least 600 mm of clear floor on the working side, so that an operator can stand without their back against a wall, and 300 mm on the other three sides for access to the isolators and the air lines. The frame itself occupies the breadboard footprint plus a small overhang, so a 750 x 900 mm bench needs roughly 1350 x 1500 mm of floor to be usable rather than merely installed.

Usage Precautions

The precautions below are the ones that matter specifically for a small frame at standing height. Most of them are about the air supply and about the three height numbers being kept apart, because those are the two things that go wrong on a bench of this kind.

  • Do not treat 800 mm as a working height. Add the breadboard thickness and the mount centreline before comparing with elbow height, and write the resulting axis height on the layout drawing so that it is not rediscovered later.
  • Keep the three adjustment figures separate. ±6 mm is the isolator travel, ±10 mm is the leg foot travel, and 26 mm (1.02 in) is the total system range. Only the last of these can be used to trim the installed height, and it is not a substitute for choosing the right frame.
  • Supply dry, oil free air at no more than 690 kPa (100 psi). Moisture and oil carry over will block the leveling valves and the damping orifices, and the symptom is a bench that drifts slowly out of level rather than one that fails outright, which is harder to diagnose.
  • Level the frame on its feet before pressurising, then pressurise and let the self-leveling settle before bolting anything down. Leveling against a pressurised frame fights the valves.
  • Distribute the load before tightening. The self-leveling system compensates for an off centre load within its range, but it cannot compensate for a load that puts one isolator outside its stroke. Place the heaviest items near the centre of the four isolators.
  • Allow the frame to settle after a large load change. The leveling valves restore position to ±0.25 mm, but the settling takes a few seconds and any measurement started during that time will drift.
  • Do not use the frame as a rigid table for a measurement that needs rigidity. With the air supply connected the surface is deliberately soft; a below 1.6 Hz corner means the bench moves under a hand push. If a procedure needs a rigid datum, remove the air and let the breadboard rest on the frame.
  • Keep the rolling diaphragm clean. The vertical seal is a rolling rubber diaphragm and it is the one part of the isolator that is sensitive to grit. Do not blow dust around the base of the isolator with compressed air.
  • If casters are fitted, do not leave the frame standing on them for a long campaign. The casters are for moving, the feet are for working, and a frame left on casters is a frame standing on four small contact patches with no leveling.
  • Check the floor. A 700 kg frame on four feet puts a substantial point load on each, and a raised access floor designed for a person and a trolley is not necessarily designed for 175 kg on a 50 mm pad. Where the floor is suspect, spread the load with plates under the feet.
  • Do not stack equipment above the bench top. The 800 mm frame raises the surface, it does not raise the ceiling, and a shelf over a standing station at 860 mm leaves very little headroom for an operator who is leaning in.

One point about the 750 x 900 mm outline deserves a separate mention. The bench is small enough that the rear 300 mm of depth is outside comfortable reach, so the layout has to be drawn with that in mind from the start. Anything that will be adjusted during a session – a kinematic mount, a stage micrometer, a fiber chuck – belongs in the front 450 mm. Items that are set once and left alone, such as a laser head, a beam dump or a fixed fold mirror, can go at the back. Reversing that arrangement is the most common reason a bench this size feels cramped even though it is not full.

The second point is about the diagonal. At 1171.5 mm the diagonal of this outline is short enough that the frame will pass through a standard 900 mm door on its side without being dismantled, provided the breadboard is removed first. That is one of the practical advantages of staying below 1000 mm on the short side, and it is worth confirming against the actual door width before ordering, because a 750 x 900 mm frame that will not go through the door is a frame that will be unloaded in a corridor.

Disclaimer

The specifications on this page are reproduced from Thorlabs published product information for the PFA series and are provided for selection and planning. Where the current official datasheet, drawing or installation manual differs from this summary, the official document takes precedence, and the manufacturer may revise specifications without notice.

The surface heights in Table 2, the axis heights in Table 3, and the reach, band and stance figures in the overview and the selection guide are calculations, not manufacturer data. They assume the breadboard thicknesses and mount centreline heights stated in the tables, a standing operator of 50th percentile Chinese adult anthropometry as described in GB/T 10000-1988, a comfortable working reach of about 65 percent of maximum forward reach, and a standing clearance of 100 mm from the front edge of the bench. Different assumptions will move the numbers, and they should be recalculated for a specific population or a specific mount.

Load capacity, resonant frequencies and leveling repeatability are quoted for the conditions defined by the manufacturer and will vary with the payload installed, its distribution across the bench, the floor and the quality of the air supply. No transmission curve for the isolation system is reproduced here; where a project requires a specific isolation performance at a specific frequency, request current performance data for the exact configuration.

No price is stated on this page. Availability, lead time and price are confirmed at the time of enquiry and may change with stock and currency. Nothing on this page constitutes a warranty, and the terms of sale are those agreed at the time of order.

Contact Us

For a quotation on PFA52505, or for help deciding whether the 800 mm frame puts the optical axis where your operators need it, send the breadboard model and thickness, the mount centreline height for the tallest optic on the bench, whether the station will be used standing or seated, the height of the people who will use it most, and confirmation that a dry air supply at up to 690 kPa is available. Those five items settle both the frame height and the isolation question.

For a project that is choosing between this outline and the 750 x 1200 mm frame, say so and include the layout sketch with the total beam path length. The two have the same depth and the same isolation, so the decision turns on whether the path fits within a 700 mm clear span, which is a question about the drawing rather than about the frame.