QQ: 859818709

Thorlabs PF10-03-F01 Ø25.4 mm UV-Enhanced Aluminum Mirror, 250-450 nm

Thorlabs PF10-03-F01 Ø25.4 mm UV-Enhanced Aluminum Mirror, 250-450 nm

Price on request
Category:

Description

Product Overview

The Thorlabs PF10-03-F01 is a Ø25.4 mm plano mirror with a UV-enhanced aluminium coating on a fused silica substrate, specified from 250 to 450 nm at 90 percent average reflectance. Behind the film the part is ordinary: 6.0 mm thick with a ±0.2 mm tolerance, λ/10 figure at 633 nm, 40-20 scratch-dig, parallelism within 3 arcmin, clear aperture 90 percent of the diameter, rear face fine ground and laser engraved with the item number. The coating is the whole story, and it is the only member of this one inch family that works below 450 nm at all.

Between 250 and 450 nm the choice of metal is not a preference, it is an elimination. Silver has its plasma edge in the deep ultraviolet: a silver mirror that is excellent at 532 nm falls away steeply below roughly 350 nm and is a poor reflector at 266 nm. Gold is worse in this region, because its interband transitions absorb everything shorter than the visible, which is exactly why a gold mirror looks yellow rather than white. That leaves aluminium. Aluminium is the only metal with useful reflectance across the near and middle ultraviolet, and it has a well known weakness: it oxidises in air, and the native oxide absorbs precisely where the ultraviolet work happens.

The UV-enhanced process addresses that weakness from both ends. The aluminium is deposited so that the oxide that forms stays thin, and it is then capped with a magnesium fluoride overcoat. Magnesium fluoride rather than the silica used on the protected coatings, because silica stops transmitting near 185 nm and its absorption climbs steeply below about 200 nm, so a silica cap would eat the margin a 250 nm mirror cannot spare; magnesium fluoride stays transparent far deeper into the ultraviolet. The overcoat also keeps the aluminium from being touched, which matters because aluminium is soft and marks if you look at it wrongly.

The damage figures are the ones that decide whether the part survives a real beam. Thorlabs rates the coating at 0.25 J/cm² at 266 nm and 0.3 J/cm² at 355 nm, both measured with 10 ns pulses at 10 Hz. Those are the two wavelengths that matter for a frequency-tripled Nd:YAG, and the fact that the 355 nm figure is slightly higher is the usual nanosecond scaling with wavelength. The band edges deserve respect too: 450 nm is a specification limit, not a soft shoulder, and a 445 nm diode is already outside the guaranteed range.

Key Features

  • 90 percent average reflectance from 250 to 450 nm: the ultraviolet band that no silver or gold coating in this catalogue covers
  • Magnesium fluoride overcoat: transparent deeper into the ultraviolet than the silica seal used on protected coatings, and hard enough to protect a soft aluminium film
  • 0.25 J/cm² at 266 nm and 0.3 J/cm² at 355 nm: 10 ns, 10 Hz ratings that match the harmonics of a tripled Nd:YAG
  • λ/10 figure at 633 nm with 40-20 scratch-dig: good enough for imaging and interferometric use, not only for beam steering
  • Parallelism within 3 arcmin and 90 percent clear aperture: a wedge small enough that the rear reflection stays close to the front one
  • Rear face fine ground and laser engraved: the part number survives in a dark enclosure where the coated face cannot be read
  • Ø25.4 mm fused silica, 6.0 mm thick: fits every standard one inch mount without an adapter

Technical Specifications

Model PF10-03-F01
Coating UV-enhanced aluminium
Overcoat Magnesium fluoride (MgF2)
Wavelength range 250 nm – 450 nm
Average reflectance Ravg 90% from 250 to 450 nm
Substrate Fused silica
Diameter Ø25.4 mm (1 in)
Diameter tolerance +0.0 mm / -0.1 mm
Thickness 6.0 mm (0.24 in)
Thickness tolerance ±0.2 mm
Surface flatness λ/10 at 633 nm
Surface quality 40-20 scratch-dig
Parallelism 3 arcmin or better
Clear aperture 90 percent of the diameter
Damage threshold, pulsed 0.25 J/cm² at 266 nm (10 ns, 10 Hz, Ø0.150 mm); 0.3 J/cm² at 355 nm (10 ns, 10 Hz, Ø0.381 mm)
Damage threshold, CW Out-of-band figures published for this coating: 300 W/cm at 1.064 µm (Ø0.044 mm); 500 W/cm at 10.6 µm (Ø0.339 mm)
Rear surface Fine ground, engraved with the item number
Source Thorlabs official series specification table

Applications

  • Third harmonic delivery: folding a 355 nm beam out of a tripler and into a working head, where silver has already lost too much reflectance to be efficient
  • Ultraviolet collimation and beam shaping: turning a collimated ultraviolet beam without introducing the chromatic focal shift a lens would add
  • Ultraviolet inspection and lithographic metrology: illumination folds and imaging folds at 365 nm or 405 nm, inside the specified band
  • Fluorescence excitation paths: steering 340 to 400 nm excitation while the emission is collected elsewhere
  • Laser marking and micromachining heads: ultraviolet folds in compact heads where the optic must also survive occasional cleaning

Selection Guide

Item Diameter Thickness When to choose it
PF03-03-F01 Ø7.0 mm 2.0 mm Miniature heads, MK7 or LMR7 class mounts
PF05-03-F01 Ø12.7 mm (0.5 in) 6.0 mm Compact paths with small ultraviolet beams
PF07-03-F01 Ø19.0 mm 6.0 mm Polaris fixed mounts, OEM ultraviolet heads
PF10-03-F01 Ø25.4 mm (1 in) 6.0 mm Standard one inch mounts, the widest choice of hardware
PF20-03-F01 Ø50.8 mm (2 in) 12.0 mm Large ultraviolet beams, converging paths
PF10-03-F01-10 Ø25.4 mm (1 in) 6.0 mm The same optic in a pack of ten, for lines and spares

Series Comparison

Coating Band Average reflectance Surface quality Role
UV-enhanced aluminium (-F01) 250 – 450 nm 90% 40-20 The only option below 450 nm
Protected aluminium (-G01) 450 nm – 20 µm 90% from 450 nm to 2 µm; 95% from 2 to 20 µm 40-20 Broadband workhorse, tolerates humidity
Protected silver (-P01) 450 nm – 20 µm 97% from 450 nm to 2 µm; 95% from 2 to 20 µm 40-20 Brightest film in the visible and near infrared
Protected gold (-M01) 800 nm – 20 µm 96% 40-20 Infrared fold, flat response
MIR-enhanced gold (-M02) 2 – 20 µm 98% 40-20 Mid-infrared fold

FAQ

Q: Why can I not just use the protected silver mirror at 355 nm?

A: You can, and it will work, but you will lose a large part of the beam at every bounce. Silver reflectance collapses below about 350 nm, so a 355 nm fold on silver wastes far more power than the same fold on UV-enhanced aluminium, and the loss is not constant across the beam. For a single diagnostic bounce it hardly matters; for a delivery path it does.

Q: Is 405 nm inside the specified band?

A: Yes, 405 nm sits comfortably inside 250 to 450 nm. The edge to watch is the long-wavelength one: 450 nm is a hard limit, reflectance falls away quickly beyond it, and a 445 nm source is already outside the guarantee even though it looks close on paper.

Q: Can the coated face be cleaned?

A: Blow it off with dry nitrogen or filtered air first, and stop there if the contamination is loose. Aluminium is soft and the fluoride overcoat is thin, so a wipe is a last resort: lens tissue, solvent, one pass, no pressure, no circular motion. Repeated wiping is the main way these coatings die.

Q: How long does the coating last under ultraviolet exposure?

A: Long enough for most industrial use, but not indefinitely. Ultraviolet irradiation and the ozone that comes with it attack the film slowly, and every cleaning cycle removes a little of the overcoat. Track hours and throughput, inspect periodically, and keep spares if the process cannot stop.

Q: Is 90 percent an average or a minimum?

A: An average across 250 to 450 nm, which is the manufacturer’s convention for metal coatings. Local values dip near the band edges and rise in the middle, so for a single wavelength near 450 nm plan for less than the average rather than exactly the average.

Q: What is the difference between this and the ten pack?

A: Nothing optical. The ten pack carries the same optic under the part number PF10-03-F01-10 and exists for lines that need spares on the shelf; buy the single unit when you need one or two.

Q: Will the rear reflection cause trouble?

A: The rear face is fine ground, so it scatters rather than forming a clean ghost beam, which is what you want. With 3 arcmin parallelism the front and rear reflections are close together, and a scattered rear lobe around the main beam is normal. Do not put a detector where that lobe lands.

Usage Precautions

  • Ultraviolet is an eye hazard even when the beam is invisible. Interlocks, shielding and appropriate goggles are part of the installation, not accessories to it.
  • Ventilate. Ultraviolet beams in air generate ozone, and in a closed enclosure the ozone attacks both the coating and anything organic nearby.
  • Do not wipe the coated face routinely. Blow off dust, and reserve solvent wipes for contamination that will not come off otherwise.
  • Respect the band edges. Below 250 nm and above 450 nm the coating is out of specification, and the reflectance you get there is not the reflectance you paid for.
  • Do not exceed the rated fluence. The 266 nm and 355 nm ratings assume 10 ns pulses at 10 Hz; shorter pulses or higher repetition rates change the number.

Disclaimer

Specifications on this page are compiled from the manufacturer’s published series tables and are provided for selection purposes only. The current official datasheet governs. Dimensions, coatings and markings may change without notice. Price and delivery are quoted separately by our sales desk.

Contact Us

Questions about fit, coating choice or lead time go to our technical desk. Send the part number, the wavelength, the beam diameter at the optic and the intended incidence angle, and you will get an answer that names a specific part rather than a range of options. Bulk and project quantities are quoted individually.

Reviews

There are no reviews yet.

Be the first to review “Thorlabs PF10-03-F01 Ø25.4 mm UV-Enhanced Aluminum Mirror, 250-450 nm”

Your email address will not be published. Required fields are marked *

中文EN