How do you choose a laser power meter sensor head? It comes down to four things: wavelength coverage, average power range, spot diameter versus effective aperture, and — for pulsed lasers — single-pulse energy and energy density checks. This article turns the official specifications of Ophir’s (an MKS company) 136 current power sensor heads into an online selection tool — enter your laser parameters and get model recommendations ranked by match score in 30 seconds, and for pulsed lasers it also runs the damage safety check automatically.
How to Use the Selector Tool
Continuous-wave (CW) lasers: enter the wavelength (nm) and the upper and lower limits of average power (W), then click “Start Selection”. The list is scored comprehensively on wavelength coverage, power margin, aperture margin, and other factors, with ⭐ marking the best recommendation. Spot diameter is optional and is used to check whether the sensor’s effective aperture leaves enough margin.
Pulsed lasers: switch to “Pulse Mode” and fill in single-pulse energy, repetition rate, pulse width, and spot diameter. The tool automatically calculates peak power, back-calculates average power from the pulse parameters and cross-checks it against the input value, computes spot area and energy density, and gives red/yellow warnings against each sensor’s 7ns / 2ms damage thresholds.
Spec-table filtering: the full specification table at the bottom of the page supports dropdown filters by detector type, aperture, absorber, cooling method, and interface, range limits on wavelength, power, effective aperture, response time, and maximum pulse energy, as well as direct search by model or part number.
Laser Power Meter Sensor Head FAQ
1. Which parameters matter first when choosing a power sensor head?
Start with wavelength range and average power range: the sensor’s spectral range must cover your laser wavelength, and the power range must cover the actual average power with a margin of more than 30%. Then check whether the effective aperture is larger than the spot diameter (the spot should not exceed 70%~80% of the aperture); for pulsed lasers, also verify single-pulse energy and energy density thresholds.
2. Thermopile or photodiode sensor head?
Thermopile sensor heads cover a wide band (typically 0.19~20μm), tolerate a large power range (mW to several kW), and are wavelength-insensitive, making them the first choice for most industrial CW laser measurements; their drawback is a slower response (typically 1~3 seconds). Photodiode (PD) sensor heads respond fast (microsecond level) with high sensitivity and can measure weak light from nW to W as well as pulsed signals, but their wavelength range is narrow (commonly 200~1100nm or 350~1100nm), and high power requires attenuating filters.
3. Is average power enough for pulsed lasers?
No. A power sensor measures average power, but a pulsed laser can damage the absorber through excessive single-pulse energy density even when the average power is within limits. You must also check peak power Ppeak = E/τ, energy density H = E/[π(d/2)²], and the sensor’s damage threshold at the corresponding pulse width (Ophir typically gives 7ns and 2ms reference points). The pulse mode of this tool performs all of these calculations automatically and gives red/yellow/green prompts.
4. What do the 7ns and 2ms energy density thresholds mean?
The damage threshold depends on pulse width: for short pulses (ns class), heat diffusion is the limiting factor and the threshold scales with √τ; for long pulses (ms class), it gradually approaches the continuous-exposure value. Ophir datasheets give two reference points, 7ns and 2ms, and the tool log-interpolates between the two to estimate the threshold at the current pulse width. This result is for engineering pre-assessment; for precise judgment, refer to the official datasheet.
5. Natural convection, fan, or water cooling?
The lower the power, the simpler the cooling: milliwatts to a few watts usually need only natural convection; tens to hundreds of watts generally use fan cooling; continuous power of a kilowatt or more generally requires water cooling. Sensors in the same series often come in different cooling versions with very different maximum measurable power, so be sure to check the complete model suffix when selecting.
6. Why are there no pyroelectric sensor heads in the tool?
This edition covers 136 thermopile and photodiode sensor heads for average power measurement. Pyroelectric sensor heads are used to directly measure pulse energy (μJ to J class); their measurement objects and readout methods are different, and a separate pulse-energy sensor selection tool will be built later.
7. Can I buy directly from the selection results?
The tool results can serve as a basis for preliminary model selection. Before ordering, we recommend final confirmation against the actual optical path (spot size, incidence angle, whether OEM/integration interfaces are needed, cable length) and the official datasheet. Soar Laser supplies the full range of Ophir power meters and sensor heads. Send us your laser parameters (wavelength, power/pulse energy, spot size, pulse width, repetition rate) and our engineers will help confirm the model and lead time. Phone/WeChat: 15927220467.



Leave a Reply